窑医疗卫生装备窑2024年2月第45卷第2期悦澡蚤灶藻泽藻酝藻凿蚤糟葬造耘择怎蚤责皂藻灶贼允燥怎则灶葬造窑灾燥造援45窑晕燥援2窑February窑2024东芝Aquilion16CT磁电阻传感器故障维修1例陶亚飞1,王延辉2*(1.联勤保障部队第980医院放射诊断科,石家庄050082;2.联勤保障部队第980医院医学工程科,石家庄050082)[摘要]阐述了东芝Aquilion系列螺旋CT相较于传统CT的优势,介绍了东芝Aquilion16CT使用过程中出现的1例磁电阻传感器故障,分析了故障产生的原因,提出了具体的解决方法,为医学工程人员维修类似故障提供了参考。[关键词]东芝Aquilion16CT;磁电阻传感器;故障维修[中国图书资料分类号]R318.6曰TH774[文献标志码]B[文章编号]1003-8868渊2024冤02-0118-03DOI院10.19745/j.1003-8868.2024040FaultrepairofToshibaAquilion16CTMRsensor:1casereportTAOYa-fei1,WANGYan-hui2*(1.DepartmentofRadiography,the980thHospitalofJointLogisticsSupportForce,Shijiazhuang050082,China;2.DepartmentofMedicalEngineering,the980thHospitalofJointLogisticsSupportForce,Shijiazhuang050082,China)AbstractTheadvantagesofToshibaAquilionseriesspiralCTweredescribedoverthetraditionalCT.OnecaseofMRsensorfaultofToshibaAquilion16CTanditscausewasintroduced,andthecountermeasurewasproposedaccordingly.Referenceswereprovidedformedicalengineerstotreatsimilarfaults.[悦澡蚤灶藻泽藻酝藻凿蚤糟葬造耘择怎蚤责皂藻灶贼允燥怎则灶葬造袁2024袁45渊2冤院118-120]KeywordsToshibaAquilion16CT;MRsensor;faultrepair0引言CT的出现对于临床诊断学具有跨时代的意义,因其具有成像敏感度高、密度分辨力高、扫描速度快等优点得以迅速发展,成为临床诊断的重要辅助工具之一。东芝Aquilion系列螺旋CT是东芝公司经过多年精心研发的高端CT,其参数指标被专家称为多层螺旋CT的金标准。该系列螺旋CT拥有多排超薄物理探测器,最小采集层厚可达0.5mm,能快速进行全身各个部位及脏器的检查[1]。其三维自适应迭代剂量降低算法首创双空间迭代,整合了智能毫安调制技术用于降低辐射剂量[2],与传统滤波反投影法相比,降低了75%的辐射剂量,是CT发展史的一个里程碑。本院于2006年购买东芝Aquilion16CT,其作为科室主力机型使用时间长达16年,日均扫描患者100余例次。本文详细介绍东芝Aquilion16CT发生的1例磁电阻传感器疑难故障,分析故障产生的原因,并提出具体的...